Thermo Scientific Nicolet Apex FTIR spectrometer
05/ 11

Spectroscopy

FTIR Spectrometer

Thermo Scientific Nicolet Apex

Fourier-transform infrared spectroscopy measures which infrared wavelengths a material absorbs. Each chemical bond has its own frequencies, so the spectrum is a fingerprint of the polymer, its cure system and its additives.

Sampling is by attenuated total reflection (ATR) with diamond, germanium and ZnSe crystals, by HATR and in transmission.

A

Key specifications

7800–350cm⁻¹

Spectral range, KBr beamsplitter

<0.25cm⁻¹

Spectral resolution

50,000:1

Signal-to-noise, guaranteed

10,000+

Reference spectra in library

Wavenumber accuracy
0.02 cm⁻¹ at 2000 cm⁻¹
Rapid scan
Up to 40 spectra/s at 16 cm⁻¹
Sampling
ATR (diamond, Ge, ZnSe), HATR, transmission
Coupling
TGA-FTIR evolved gas analysis

Performance data from Thermo Fisher Scientific Nicolet Apex specifications; accessories as installed.

B

Typical applications

  • Polymer family identification (FKM, FFKM, HNBR, EPDM, PTFE, PEEK)
  • Verification of incoming material against reference spectra
  • Cure system and additive identification
  • Chemical change after fluid or thermal exposure
  • Oxidation and degradation of elastomers
  • Surface contamination and residues
  • Plasticizer and oil migration
  • Composite matrix identification
  • Reverse engineering of unknown seals
  • Evolved gas analysis coupled to TGA
C

Example outputs

FTIR spectrum of HNBR

1

HNBR spectrum

The sharp nitrile band near 2236 cm⁻¹ marks the acrylonitrile in the polymer.

FTIR spectrum of FKM

2

FKM spectrum

Strong C–F absorptions between 1100 and 1300 cm⁻¹ dominate the spectrum of a fluoroelastomer.